Six-Cup Insulating Oil Dielectric Strength Tester
This model of insulating oil dielectric strength tester is a high-precision, fully digital industrial instrument meticulously developed by our R&D team. It is the result of leveraging our technical strengths and conducting extensive field testing and long-term, dedicated effort, all in strict accordance with relevant standards and regulations for insulating oil testing. The instrument features a user-friendly interface and an elegant, professional design. Thanks to its fully automatic digital microprocessor control, it offers high measurement accuracy, strong anti-interference capabilities, and reliable safety performance.
Instrument Features:
1. Controlled by a high-capacity microcontroller for stable and reliable operation;
2. Features displays for temperature, humidity, and time, with an option for infrared oil temperature measurement. 3. Built-in wide-range watchdog circuit eliminates system freezes;
4. Multiple standard options: the instrument includes programs for GB/T507-1986, GB/T507-2002, DL429.9, and IEC156, as well as a user-programmable mode to meet diverse user requirements;
5. The oil cup is precision-manufactured using special glass and polymer materials, preventing oil leakage and corrosion;
6. Unique high-voltage side sampling design feeds test values directly into the A/D converter, avoiding errors associated with analog circuitry and ensuring greater measurement accuracy;
7. Features internal protections against overcurrent, overvoltage, and short circuits, along with strong anti-interference capabilities and excellent electromagnetic compatibility;
8. Switchable between Chinese and English interfaces to meet various customer needs;
9. Built-in data analysis function assists personnel in evaluating oil quality;
10. USB data transmission (optional);
Technical Parameters
1. Step-up transformer capacity: 1.5 kVA
2. Voltage rise rate: 0.5 kV/s–5.0 kV/s (selectable in 0.5 kV/s increments across 10 levels); error: 0.2 kV/s
3. Output voltage: 0–80 kV or 100 kV (optional)
4. Voltage accuracy: ±(2% of reading + 2 digits)
5. Power supply distortion rate: <1%
6. Electrode gap: Standard 2.5 mm
7. Number of tests: 6 (selectable range: 1–9)
8. Initial resting time: 900 s (selectable range: 0–999 s)
9. Interval resting time: 300 s (selectable range: 0–999 s)
10. Stirring time: 15 s (selectable range: 0–255 s)
11. Dimensions: 800 mm × 660 mm × 710 mm
12. Weight: 75 kg
Operating Conditions
1. Ambient temperature 0–40°C
2. Relative humidity ≤85%
3. Power supply AC 220V (1 ± 10%)
4. Power frequency 50 Hz (1 ± 10%)
5. Power consumption <200 W